Early catastrophic failure of a PINNACLE MARATHON® polyethylene acetabular liner: a report of a malpositioned PINNACLE DUOFIX HA® acetabular shell resulting in point loading and accelerated wear

Abstract

Background

Highly cross-linked polyethylene in total hip arthroplasty (THA) has been shown to decrease wear rate compared with a conventional polyethylene liner. However, it has also been reported that the manufacturing processes can cause early failure of the implant.

Case

We describe early catastrophic failure at <4 years following primary THA surgery of a PINNACLE MARATHON® polyethylene acetabular liner resulting from a malaligned PINNACLE DUOFIX HA® acetabular shell with resultant point loading. Radiographs revealed a malaligned acetabular shell and superior subluxation of the femoral head out of the liner. At revision surgery operative findings revealed that the acetabular shell alignment was in of 50° of anteversion and 70° of inclination. Significant metallosis in the surrounding tissues was observed.

Conclusions

We conclude that a malpositioned acetabular shell resulting in point loading and abnormal contact stresses was the mechanism of failure. Our case highlights the importance of achieving correct acetabular component alignment in total hip arthroplasty.

Hip Int 2016; 26(6): e52 - e55

Article Type: CASE REPORT

DOI:10.5301/hipint.5000447

Authors

Daniel S. Hill, Kuntal Patel, Deepak Herlekar

Article History

Disclosures

Financial support: None.
Conflict of interest: None.

This article is available as full text PDF.

  • If you are a Subscriber, please log in now.

  • Article price: Eur 36,00
  • You will be granted access to the article for 72 hours and you will be able to download any format (PDF or ePUB). The article will be available in your login area under "My PayPerView". You will need to register a new account (unless you already own an account with this journal), and you will be guided through our online shop. Online purchases are paid by Credit Card through PayPal.
  • If you are not a Subscriber you may:
  • Subscribe to this journal
  • Unlimited access to all our archives, 24 hour a day, every day of the week.

Authors

Affiliations

  • Department of Orthopaedic Surgery, Royal Lancaster Infirmary, Lancaster - UK

Article usage statistics