Advertisement

Evaluation of repair of the hip labrum under simulated full weight-bearing

Evaluation of repair of the hip labrum under simulated full weight-bearing

Hip Int 2017; 27(1): 104 - 109

Article Type: SURGICAL TECHNIQUE

DOI:10.5301/hipint.5000472

OPEN ACCESS ARTICLE

Authors

Jason L. Koh, Kavish Gupta

Abstract

Introduction

Repairs of labral tears are performed for unstable tears, hip instability, and after detachment concomitant to the treatment of femoroacetabular impingement (FAI), but limited data is known about the strength of repair. This study evaluated the effect of simulated axial weight-bearing on suture anchor based repair of the acetabular labrum.

Methods

3 cadaveric pelvises underwent creation of a 1.5 cm anterior-superior labral tear in each hip. The tears were then repaired using 2 suture anchors per hip. Following repair, the hip joint underwent axial cyclic loading to 756 N, and were inspected for separation of the labrum from the acetabulum. The strength of the suture anchor repair was evaluated by testing load to failure, in-line with insertion.

Results

Upon visual examination, all 6 repairs remained fully intact following loading with no visible gap formation or damage at the repair site. In all cases an arthroscopic probe could not be inserted under the edge of the repair. The mean failure force of the 12 suture anchors, in-line with insertion, was 154 N ± 44 N.

Conclusions

Acetabular labral suture anchor repairs may be able to immediately withstand the physiological loads of axial weight-bearing. Labral repair may be able to tolerate axial weight-bearing immediately after repair, preserving the strength and integrity of muscles and soft tissues.

Article History

Disclosures

Financial support: Specimens and equipment were provided by Arthrex.
Conflict of interest: JLK has received funding from Aesculap, Aperion and Arthrex.

This article is available as full text PDF.

Download any of the following attachments:

Introduction

Pathological tears of the acetabular labrum have been increasingly recognised as a cause of pain, dysfunction, and instability (1-2-3). In addition, there is increasing evidence that damage to the labrum is related to the development of osteoarthritis of the hip joint (4-5-6). The hip labrum appears to primarily function to seal the edge of the joint and has a relatively small role in load bearing (7-8-9-10). Loss of the sealing function of the labrum can lead to an increased rate of consolidation of the articular cartilage, resulting in increased contact stresses between the femoral and acetabular cartilage layers, and in some cases progression to arthritis (5, 6, 8).

Recent studies have demonstrated the favourability of labral repair when indicated (6, 11). Repair of the labrum is performed for unstable tears, hip instability, and after detachment concomitant to the treatment of femoroacetabular impingement (FAI) (11, 12). Restoration of the normal seal may slow or prevent the degenerative delamination of adjacent articular cartilage.

Initially, repair of the acetabular labrum was performed with the use of drill holes placed in the acetabular rim, similar to the original repairs of shoulder labral lesions (13). The use of tack-type or suture anchor repairs has been previously described (14). However, there is very little information regarding the mechanical strength of hip labral repairs. Therefore, it remains unclear whether these repairs can sustain the forces generated by full weight-bearing. We hypothesise that suture anchor repairs of the acetabular labrum have sufficient strength to withstand full weight-bearing due to the labrum’s limited role in load-bearing. Increased information about the mechanical strength of hip labral repairs may help to better define postoperative guidelines and rehabilitation protocol.

The purpose of this study was to test the strength of suture anchor based repair of the acetabular labrum under conditions simulating a normal weight-bearing load.

Materials and methods

Specimen preparation and alignment

6 macroscopically normal lower extremity specimens were obtained from 3 donors (25-year-old male, 52-year-old male and 62-year-old male). Specimens with osteoarthritis or obvious abnormal morphology were excluded. Muscular soft tissues were removed, leaving the joint capsule, ligaments, and labrum intact.

Specimens were mounted to simulate axial weight-bearing in neutral extension (10). This was performed as follows. Both femoral shafts of each specimen were potted in 3” schedule 40 PVC pipe using methyl methacrylate. A threaded stud was cemented in the distal end of the PVC pot. The femur was potted to allow loading of the hip in-line with the mechanical axis of the femur while in a neutral position (Line A in Figure 1). To facilitate alignment during potting, the pelvis was placed so that the anterior side of the pelvis lay on the table. The pelvis should rest on the anterior superior iliac spine and the pubic tubercle in this position. Anatomically these landmarks lie in a plane parallel to the frontal plane. A straight edge was used to create a line tangent to the right and left iliac crest (Line B in Figure 1). A line parallel to the longitudinal axis of the femur was drawn on the posterior side of the femur (Line C in Figure 1). In a neutral position the femur is adducted approximately 6°; therefore, a line 6° to Line C was created on the femoral shaft using a goniometer (Line D in Figure 1). Neutral position should be replicated when the 6° line (Line D in Figure 1) is perpendicular to Line B. During potting, the femur was positioned in the PVC pipe so that the longitudinal axis of the PVC pipe was parallel to the 6° line (Line D in Figure 1). In addition, the longitudinal axis of the pipe was positioned in-line with the mechanical axis of the femur (Line A, a line through the centre of curvature of the femoral head perpendicular to Line B).

Schematic diagram describes a posterior view of the pelvis and femur in the neutral position tested in the study.

A flat spot was created on both sides of the iliac fossa using a PVC pot and bone cement to facilitate fixation of the pelvis in the INSTRON machine. The flat spot was created in-line with the mechanical axis of the femur in the neutral position as seen in Figure 1.

Creation of labral tears

Labral tears were created and repaired after potting to reduce unnecessary loading to the repairs during the potting process. The anterior-superior rim is the most common location for acetabular tears (15) and is the area reattached in labral repair for injury or in the treatment of femoroacetabular impingement. To simulate these tears, the joint capsule was removed and 1.5 cm of the anterior-superior labrum was sharply detached longitudinally from the acetabular rim, using a scalpel as seen in Figure 2A. If a small partial thickness tear was identified, the labral tear was completed to full thickness and then repaired.

(A) Photograph shows the anterior-superior labrum detached from acetabular rim using a scalpel; (B) Photograph shows a labral repair using Bio-PushLocks; (C) Photograph shows a labral repair using Bio-SutureTaks.

Labral repair

Suture anchors were used to fixate the labrum to the acetabular rim. Labral repairs were performed by placing two suture anchors 5 mm apart, dividing the area of detachment into thirds. Matched pairs of repairs, using either 3.0 mm BioComposite Bio-SutureTaks or 3.5 mm Bio-PushLocks (Arthrex, Inc.), were performed in left and right hips of each specimen (FDA-approved for this use). Each specimen received a repair using each type of anchor. #2 FiberWire (Arthrex, Inc.) was used in both suture anchor groups, passed circumferentially around the tear. The tied sutures were secured with alternating half hitches. Figure 2 illustrates labral repair using Bio-PushLocks (2B) or Bio-SutureTaks (2C).

Securing and aligning the specimen

The pelvis and femur were secured in an inverted position in an Instron 8871 materials testing machine (Instron). The flat spot created on the iliac was clamped in an adjustable angle fixture so that the anterior-superior iliac spine and the pubic tubercle were all in a plane perpendicular to the base of the Instron machine (Fig. 3). The pelvis was aligned to be in neutral tilt in the coronal and saggital planes using a carpenter’s bubble level. The right and left ischial tuberosity were level relative to each other. The potted femur was attached to the load cell by a vise in 12° of anteversion and 7° of varus with the mechanical axis aligned with the femoral head. The overall alignment was designed to simulate single-leg stance, which was chosen because of its extensive utilisation during daily activities such as walking and climbing stairs.

Photograph demonstrates a pelvis and femur mounted in the INSTRON machine in the inverted position tested in the study.

Mechanical testing

The femur was axially loaded to simulate loads placed on the hip joint during activity (10). The hip was preloaded to 132 N (the approximate load during toe-touch weight-bearing). Cyclic loading was conducted from 132 N to 756 N at 1 Hz, for 100 cycles. 756 N was chosen to simulate full weight-bearing of an average adult male (77 kg). Full weight-bearing would not be recommended during rehabilitation; however, it was tested to simulate worse-case loading scenarios. Post cycling, the joint was removed from the frame and the femoral head was distracted from the acetabulum and the labral repair was examined. Using a pair of digital calipers, gap displacement of the labrum relative to the acetabular rim was measured.

Following examination of the labral repair, pull to failure of the suture anchors was conducted. The flat spot created on the iliac of the pelvis was clamped in a vise-clamp fixated to an adjustable angle fixture. The angle of the vise-clamp was adjusted so that pull to failure in-line with the insertion of the suture anchors could be conducted. In-line axial pull of the suture would not be a physiological load applied to suture anchors used for labral repairs; however, it was chosen to test a worse-case scenario. The free ends of the suture strands were clamped in a pneumatic clamp specifically designed for fixating suture. Pull to failure was conducted at a constant displacement rate of 12 in/min. Data was collected at 500 Hz.

Results

All specimens had no gross evidence of osteoarthritis. 4 hips had small partial thickness tears at the articular side of the anterior-superior labrum, consistent with labral tears caused by cam-type femoroacetabular impingement. During examination of the acetabulum of specimen 05 08014, a 1.6 mm pre-existing tear was found in the labrum (Fig. 4A). The pre-existing tear was repaired using Bio-SutureTaks. In this specimen some delamination of the lunate cartilage was present inside the acetabulum (Fig. 4A). 2 hips had minimal changes at the femoral head-neck junction and 2 hips had a bony prominence at the femoral head-neck junction. All hips had a normal centre-edge angle (>25°).

(A) Photograph shows a specimen with pre-existing tear and associated articular cartilage lesion after repair and cycling; (B) Photograph shows a labral repair after cycling: labrum repaired with Bio-PushLocks; (C) Photograph shows a labral repair after cycling: labrum repaired with Bio-SutureTaks.

Following the creation of the 1.5 cm detachment, the mean amount of displacement of the labrum from the edge of the acetabulum was 3 mm.

Visual inspection following cyclic loading demonstrated no obvious changes at the repair site (Fig. 4B, C). In all cases, an arthroscopic probe (diameter 1 mm) could not be placed under the edge of the repair.

The average pull-out force for all 12 suture anchors from the acetabular bone was 154 N ± 44 N. Bio-SutureTaks had a significantly (p<0.009) higher pull-out strength than Bio-PushLocks (180 N ± 36 N vs. 129 N ± 38 N). No significant difference between anterior and posterior anchor pull-out was noted (153 N vs. 156 N). The data for each individual specimen can be seen in Table IA, B.

Raw data for anchor pull-out from the acetabular rim

TABLE I A - Bio-PushLock
Donor Location Ultimate Force (N) Mode of Failure
05 05059 Anterior 103 Anchor pull-out
05 05059 Posterior 97 Anchor pull-out
05 08014 Anterior 183 PEEK tip pull-out
05 08014 Posterior 170 Anchor pull-out
05 07002 Anterior 127 Anchor pull-out
05 07002 Posterior 96 Anchor pull-out
Number (n) 6
Average 129
Standard Deviation 38
TABLE I B - Bio-SutureTak
Donor Location Ultimate Force (N) Mode of Failure
05 05059 Anterior 154 Eyelet break
05 05059 Posterior 154 Eyelet break
05 08014 Anterior 180 Eyelet break
05 08014 Posterior 249 #2 FiberWire Broke
05 07002 Anterior 169 Eyelet break
05 07002 Posterior 171 Eyelet break
Number (n) 6
Average 180
Standard Deviation 36

Discussion

There is increasing interest in the treatment of labral pathology of the hip. It is clear that damage to the labrum can cause pain and loss of function. The loss of the sealing effect of the joint may also result in the development of articular cartilage damage and contribute to accelerated osteoarthritis of the hip.

Most labral tear lesions likely occur as the result of femoroacetabular impingement (FAI), which is the repetitive compression from the femoral head and neck on the anterior/anterior-superior labrum. This type of impingement typically occurs with the hip in >90° of flexion and internal rotation, but may occur if there are morphological abnormalities of the femoral head-neck junction. In several of our specimens, the combination of a small labral tear, articular cartilage damage and impinging bone at the femoral head-neck junction was seen.

Although the labrum adds to the area of the hip joint socket (16), it is unclear what loads are placed on the anterior-superior labrum during single leg stance or the gait cycle (9, 10) Konrath et al (10) concluded that “In contrast to the meniscus of the knee, the acetabular labrum does not increase contact area, distribute load, or reduce contact stresses in the hip.” Henak et al (9) found that the acetabular labrum supports only 1-2% of the total load placed on the hip during daily activities. Safran et al (17) determined that the greatest strain in the intact labrum in full extension is laterally and strain is highest in the anterior labrum in flexion and adduction. However, the overall magnitude of average strain in the labrum was extremely small (<0.04 mm). Thus, our results are consistent with Konrath et al (10), Henak et al (9), and Safran et al (17), in that no changes at the repair site were seen despite the repeated loads applied to the hip joint in our described axial positioning. This suggests that labral repairs may have sufficient initial strength to withstand loads associated with axial weight-bearing.

We recognise that this study did not evaluate the hip through a range of motion. Data from Henak et al (9) and Safran et al (17) suggest that magnitude strain in the labrum under higher degrees of flexion is quite low, suggesting that these results are applicable through a relatively broad range. The highest negative displacement of the normal labrum in 90° of hip flexion with abduction and internal rotation was 0.014435 mm. Axial loading to simulate body weight was performed due to recognition that changing the position was unlikely to change the result, given the findings of Safran et al (17).

During these normal activities, the labrum appears to be only minimally loaded. The postoperative recommendations following labral repair are therefore relatively precautionary, with respect to protecting the repair. Typically, they require limited weight-bearing for 4-8 weeks, limited rotation, and no flexion >70° (18). Patients are advised to avoid placing excessive load on the repair until biological healing of the damaged tissue can occur. The findings in this study suggest that post-operative weight-bearing may be able to be progressed earlier.

Increased postoperative weight-bearing could help patients maintain normal gait patterns as well as preserve the strength and integrity of muscles and other soft tissues. Further clinical investigation is necessary to determine whether altering rehabilitation protocol and postoperative recommendations is advisable.

We feel that suture anchor-based repair of acetabular labral tears has sufficient initial strength to withstand weight-bearing loads in the postoperative period. It is important that if the mechanism of injury is felt to be femoroacetabular impingement, it be concurrently treated either with open or arthroscopic procedure, otherwise the potential for re-tear exists (19). In the dysplastic hip, a pelvic osteotomy to redistribute the load to the acetabulum may be necessary (15, 20).

In addition, the loads on the labrum of a dysplastic hip are likely to be substantially higher than in the normal joint, since the normal load transmission through the acetabular roof is significantly decreased (9). This study was not designed to evaluate this condition. It is possible that in a dysplastic hip, greater load transmission across the repair would occur, leading to gap formation or fracture of the repair.

The weaknesses of this study include the fact that it was a cadaveric study and confined to anatomically normal hips. Also, the decision was made to use axial loading, which did not simulate the full range of gait. We feel that this study still demonstrates the effects of early weight-bearing on acetabular labral repair and may provide a guide on postoperative and rehabilitation protocols.

Conclusion

Acetabular labral suture anchor repairs may be able to immediately withstand the physiological loads of axial weight-bearing. Labral repair may be able to tolerate axial weight-bearing immediately after repair, preserving the strength and integrity of muscles and soft tissues. Further clinical investigation is needed to determine whether postoperative weight-bearing is advisable immediately following acetabular labral suture anchor repairs.

Disclosures

Financial support: Specimens and equipment were provided by Arthrex.
Conflict of interest: JLK has received funding from Aesculap, Aperion and Arthrex.
References
  • 1. Burnett RS Della Rocca GJ Prather H Curry M Maloney WJ Clohisy JC Clinical presentation of patients with tears of the acetabular labrum. J Bone Joint Surg Am 2006 88 7 1448 1457
  • 2. Nepple JJ Philippon MJ Campbell KJ et al. The hip fluid sealPart II: The effect of an acetabular labral tear, repair, resection, and reconstruction on hip stability to distraction. Knee Surg Sports Traumatol Arthrosc 2014 22 4 730 736
  • 3. Smith MV Panchal HB Ruberte Thiele RA Sekiya JK Effect of acetabular labrum tears on hip stability and labral strain in a joint compression model. Am J Sports Med 2011 39 1_suppl Suppl 103S 110S
  • 4. Jessel RH Zurakowski D Zilkens C Burstein D Gray ML Kim YJ Radiographic and patient factors associated with pre-radiographic osteoarthritis in hip dysplasia. J Bone Joint Surg Am 2009 91 5 1120 1129
  • 5. McCarthy JC Noble PC Schuck MR Wright J Lee J The Otto E. Aufranc Award: The role of labral lesions to development of early degenerative hip disease. Clin Orthop Relat Res 2001 393 25 37
  • 6. Song Y Ito H Kourtis L Safran MR Carter DR Giori NJ Articular cartilage friction increases in hip joints after the removal of acetabular labrum. J Biomech 2012 45 3 524 530
  • 7. Dwyer MK Jones HL Hogan MG Field RE McCarthy JC Noble PC The acetabular labrum regulates fluid circulation of the hip joint during functional activities. Am J Sports Med 2014 42 4 812 819
  • 8. Ferguson SJ Bryant JT Ganz R Ito K An in vitro investigation of the acetabular labral seal in hip joint mechanics. J Biomech 2003 36 2 171 178
  • 9. Henak CR Ellis BJ Harris MD Anderson AE Peters CL Weiss JA Role of the acetabular labrum in load support across the hip joint. J Biomech 2011 44 12 2201 2206
  • 10. Konrath GA Hamel AJ Olson SA Bay B Sharkey NA The role of the acetabular labrum and the transverse acetabular ligament in load transmission in the hip. J Bone Joint Surg Am 1998 80 12 1781 1788
  • 11. Larson CM Giveans MR Stone RM Arthroscopic debridement versus refixation of the acetabular labrum associated with femoroacetabular impingement: mean 3.5-year follow-up. Am J Sports Med 2012 40 5 1015 1021
  • 12. Espinosa N Beck M Rothenfluh DA Ganz R Leunig M Treatment of femoro-acetabular impingement: preliminary results of labral refixation. Surgical technique. J Bone Joint Surg Am 2007 89 Suppl 2 Pt.1 36 53
  • 13. Lieberman JR Altchek DW Salvati EA Recurrent dislocation of a hip with a labral lesion: treatment with a modified Bankart-type repair. Case report. J Bone Joint Surg Am 1993 75 10 1524 1527
  • 14. Mei-Dan O Pascual-Garrido C Kark JA McConkey MO Inside out: a novel labral repair and advancement technique. Arthrosc Tech 2014 3 2 e241 e244
  • 15. Ross JR Zaltz I Nepple JJ Schoenecker PL Clohisy JC Arthroscopic disease classification and interventions as an adjunct in the treatment of acetabular dysplasia. Am J Sports Med 2011 39 1_suppl Suppl 72S 78S
  • 16. Tan V Seldes RM Katz MA Freedhand AM Klimkiewicz JJ Fitzgerald RH Jr Contribution of acetabular labrum to articulating surface area and femoral head coverage in adult hip joints: an anatomic study in cadavera. Am J Orthop (Belle Mead NJ) 2001 30 11 809 812
  • 17. Safran MR Giordano G Lindsey DP et al. Strains across the acetabular labrum during hip motion: a cadaveric model. Am J Sports Med 2011 39 1_suppl Suppl 92S 102S
  • 18. Lewis CL Sahrmann SA Acetabular labral tears. Phys Ther 2006 86 1 110 121
  • 19. Larson CM Giveans MR Samuelson KM Stone RM Bedi A Arthroscopic Hip Revision Surgery for Residual Femoroacetabular Impingement (FAI): Surgical Outcomes Compared With a Matched Cohort After Primary Arthroscopic FAI Correction. Am J Sports Med 2014 42 8 1785 1790
  • 20. Kim KI Cho YJ Ramteke AA Yoo MC Peri-acetabular rotational osteotomy with concomitant hip arthroscopy for treatment of hip dysplasia. J Bone Joint Surg Br 2011 93 6 732 737

Authors

Affiliations

  • NorthShore Orthopaedic Institute, NorthShore University HealthSystem, Evanston, Illinois - USA

Article usage statistics

The blue line displays unique views in the time frame indicated.
The yellow line displays unique downloads.
Views and downloads are counted only once per session.

No supplementary material is available for this article.